Product Code: SYK-1C, Versatile Anesthesia Machine
Main Specifications:
1.Flow:O2 0.1~15L/min
N2O 0.05~10L/min
ORC O2≥25%
2.O2 Flush: 35~75L/min
3.Oxygen deficiency alarm and N2O cut-off
when O2≤0.2Mpa,audible alarm occurs and lasts for at least 7 seconds
when O2≤ 0.14Mpa,N2O flow decreases correspondingly
when O2≤ 0.04Mpa,N2O flow is cut off
4.Integrated breathing circle system
APL adjustment range:0.5~7kPa
CO2 absorber acpcity:1.6L
Vaporizer(Drager19.3)
Ambient temperature 15~35℃
Input flow 0.25~15L/min
Output back pressure:-10kPa~20kPa
Output concentration:0.2~5Vol%±20%(OR±0.2%)
Anesthesia Ventilator:
A microprocessor-controlled ventilator,which is time-cycled and constant-volume,has a clear digital reading for all set breathing parameters.Because it is electric-driven and electric-controlled,the ventilator saves large amount of oxygen that is usually used only as driving gas.
Monitor:
Respiratory monitor applies microprocessor to monitor Tidal Volume,Frequency,MV airway pressure,and inhaled oxygen concentration.It has real time curve display and a visual-and-audible alarms.A Chinese menu and single-key operation is userfriendly. A high-definition large screen with wide vision angle make all-directional inspection available.
Specification:
Frequency:6-60bmp
O2 measuring range 0~99%
Tidal Volume:100~1500mL
Minute volume:0-40L/min
Airway pressuer:0-9.9kPa
M-IC is a new generation anaesthesia machine incorporating anaesthesia,respiratory and monitoring functions.With the adoption of modem art and somatological engineering in its unique designation,the machine is not only a state of art but also of great convenience in manipulation.
Anesthesia Machine:
It adopts a two-stage pressure regulator and a high precision dual-tube flowmeter to provide an accurate fresh gas output.Its build-in safety devices such as ORC system, oxygen deficiency alarm and N20 cut-off make anaesthesia more safe. The specially constructed inter-lock mechanism may guarantee only one of the two parallel vaporizers to be switched on at a time.The selected vaporizers from world famous companies as Drager 19.3, Ohmeda TEC-5 and Penlon Sigma Delta satisfy different requirements of customers.
With SYK-1C it is possible to act low-flow anaesthesia because of its unique integrated circle system-a system that keeps the inhaled tidal volume from fresh gas. So it not only saves anaesthesia agent but also reduces pollution to operation room.